Identifying ex vivo acute ischemic stroke thrombus composition using electrochemical impedance spectroscopy

Identifying ex vivo acute ischemic stroke thrombus composition using electrochemical impedance spectroscopy

Background

Intra-procedural characterization of stroke thromboemboli might guide mechanical thrombectomy (MT) device choice to improve recanalization rates. Electrochemical impedance spectroscopy (EIS) has been used to characterize various biological tissues in real time but has not been used in thrombus.

Objective

To perform a feasibility study of EIS analysis of thrombi retrieved by MT to evaluate: (1) the ability of EIS and machine learning to predict red blood cell (RBC) percentage content of thrombi and (2) to classify the thrombi as “RBC-rich” or “RBC-poor” based on a range of cutoff values of RBC.

Methods

ClotbasePilot was a multicentric, international, prospective feasibility study. Retrieved thrombi underwent histological analysis to identify proportions of RBC and other components. EIS results were analyzed with machine learning. Linear regression was used to evaluate the correlation between the histology and EIS. Sensitivity and specificity of the model to classify the thrombus as RBC-rich or RBC-poor were also evaluated.

Results

Among 514 MT,179 thrombi were included for EIS and histological analysis. The mean composition in RBC of the thrombi was 36% ± 24. Good correlation between the impedance-based prediction and histology was achieved (slope of 0.9, R2  =  0.53, Pearson coefficient  =  0.72). Depending on the chosen cutoff, ranging from 20 to 60% of RBC, the calculated sensitivity for classification of thrombi ranged from 77 to 85% and the specificity from 72 to 88%.

Conclusion

Combination of EIS and machine learning can reliably predict the RBC composition of retrieved ex vivo AIS thrombi and then classify them into groups according to their RBC composition with good sensitivity and specificity.
Identifying ex vivo acute ischemic stroke thrombus composition using electrochemical impedance spectroscopy                      Identifying ex vivo acute ischemic stroke thrombus composition using electrochemical impedance spectroscopy
Identifying ex vivo acute ischemic stroke thrombus composition using electrochemical impedance spectroscopy

Learn more about the ISX-3 impedance analyzer

Sciospec ISX-3
Class-Leading Impedance analyzer, the Sciospec ISX-3

Get to know the ISX-3 impedance analyzer

Latest Products

  • Sciospec EIT-lung imaging kit for impedance tomography
    LungEIT Kit
    0.00

    Das Sciospec LungEIT Kit ist die schlüsselfertige Lösunge für Lungenbildgebung mit elektrischer Impedanztomograpahie.

    Learn more
  • Sciospec LCR-1
    LCR-1
    0.00

    kompaktes LCR-Meter mit riesigem Messbereich und intuitiver Bedienung

    Learn more
  • Medical Research ISX-3
    0.00

    single channel impedance analyzer with medical grade safety features

    Learn more

Get in touch with us

+49 3425 88399 00

info@www.sciospec.com

Sciospec Scientific Instruments GmbH
Leipziger Str. 43b, 04828 Bennewitz, Germany