Measuring the viscosity of glycerin solutions in very small samples using Sciospec ISX-3 Impedance Analyzer

Measuring the viscosity of glycerin solutions in very small samples using Sciospec ISX-3 Impedance Analyzer

Sciospec ISX-3 impedance analyzerMeasuring the viscosity of very small sample volumes is of high interest in many fields especially in miniaturized and micro-fluidic systems. This application note describes an approach using the change of the impedance spectrum of a Thickness Shear Mode Resonator in combination with a sophisticated fitting algorithm. The simple setup of the measuring system enables easy integration in existing micro-fluidic settings with almost no adaption. As validation of the Sciospec ISX-3 a comparison to theoretical values and to a commercially available vector network analyzer on the same setup is provided.

The measurement of the viscosity is very well known for sample volumes larger than a couple milliliters. The standard viscometers use either rotating parts, a flow through, a capillary system or falling spheres. This application note describes an example of measuring smaller sample volumes with no need for mechanical parts.

 

 

 

Setup

A thickness shear mode resonator (TSM) with a nominal frequency of 9 MHz was connected to the Sciospec ISX-3 by a custom adapter with SMA connectors. Since the sensitivity is highly dependent on the temperature of the sensor chip it was held constant at 25°C with an external temperature controller. The frequency range was set to 8.9 MHz to 9.1 MHz with 1001 linear distributed points. The inter-frequency delay was set to 10 ms, so the quartz system was allowed enough time to adapt to the change in generator frequency. As control fluid a sample of 1.2 ml double distilled Water was added to the sensor. After 5 min of equalization of temperature the impedance spectrum was measured and saved with the Sciospec Impedance Analyzer suite. After thorough rinsing of the TSM chip, this procedure was repeated with 10, 30, 50 and 70 mass-% glycerin solution. All impedance spectra are shown in Fig 1.

Measuring the viscosity of glycerin solutions in very small samples using Sciospec ISX-3 Impedance Analyzer

Evaluation

The resulting impedance spectra were fitted to a complex model2 developed by the Surface Dynamics Group of the Leibnitz Institute for Solid State and Materials Research Dresden. The model not only takes advantage of the main mode of the oscillator it also includes secondary modes. A total number of 5 modes per resonance were analyzed to estimate the viscosity with high precision (±5 %). This allows for a stable estimation since the whole impedance spectrum is used and not only characteristic frequencies or impedance values. Even a multiple layer  system with estimation of the viscosity or density of each layer itself can be analyzed with this method. In this study a one inch quartz crystal was used.

Measuring the viscosity of glycerin solutions in very small samples using Sciospec ISX-3 Impedance Analyzer

The sample volume of 1.2 ml can be further reducedthrough the integration of smaller TSM sensor chips. As seen in Tab 1 and Fig 2 the resulting values for the viscosity are in good agreement with the theoretical values taken from “Physical properties of glycerine and its solutions” which were obtained by measuring much larger sample volumes.

Measuring the viscosity of glycerin solutions in very small samples using Sciospec ISX-3 Impedance Analyzer

Additional Validation

To validate the accuracy of the Sciospec ISX-3 system two methods were used. First of all a statistical analysis of each measurement was done to show the reproducibility of the system setup. Secondly a comparison to a  conventional test setup was done by measuring the exact same sensor with an Agilent vector network analyzer E5071C. Both test results match to a very high degree and conclude in the same viscosity estimation. Fig 3 as an example shows the impedance spectra for the 30 mass-% glycerin – water solution measured with both devices. Almost no difference in the impedance spectra was found.

Measuring the viscosity of glycerin solutions in very small samples using Sciospec ISX-3 Impedance Analyzer

Outlook

The experiments proofed that Sciospec measuring technology compares well to conventional setups for the use with high-Q resonant sensors for analysis of ultra small fluid samples. Since Sciospec devices are highly modular, adaption to more complex application requirements is easily done. One module already in use integrates a high precision temperature controller into the instrument enabling complete standalone solution for this type of experiment. Multichannel solutions are currently in development. A novel algorithm for Sciospec devices reduces the interfrequency delay for resonant systems enabling ultra fast measurements. The complex model approach describe above will include multilayer parameter estimation of layer forming fluids.

Measuring the viscosity of glycerin solutions in very small samples using Sciospec ISX-3 Impedance AnalyzerThis application note is part of the results of the joint research between Sciospec and the Leibniz Institute for solid State and Materials Research (IFW) Dresden. The project “Technologieplattform: Akustoelektronische Mikrofluidik” was funded by the BMBF program InnoProfile.

Latest Products

  • Sciospec EIT-lung imaging kit for impedance tomography
    LungEIT Kit

    The Sciospec LungEIT Kit is your turn-key, easy entry into lung imaging with impedance tomography.

    Learn more
  • Sciospec LCR-1
    LCR Meter

    powerful, yet easy to use LCR meter

    Learn more
  • Medical Research ISX-3

    scalable impedance analyzer with medical grade safety features

    Learn more

Get in touch with us

+49 3425 88399 00

info@www.sciospec.com

Sciospec Scientific Instruments GmbH
Leipziger Str. 43b, 04828 Bennewitz, Germany