Measuring the viscosity of glycerin solutions in very small samples using Sciospec ISX-3 Impedance Analyzer
Measuring the viscosity of very small sample volumes is of high interest in many fields especially in miniaturized and micro-fluidic systems. This application note describes an approach using the change of the impedance spectrum of a Thickness Shear Mode Resonator in combination with a sophisticated fitting algorithm. The simple setup of the measuring system enables easy integration in existing micro-fluidic settings with almost no adaption. As validation of the Sciospec ISX-3 a comparison to theoretical values and to a commercially available vector network analyzer on the same setup is provided.
The measurement of the viscosity is very well known for sample volumes larger than a couple milliliters. The standard viscometers use either rotating parts, a flow through, a capillary system or falling spheres. This application note describes an example of measuring smaller sample volumes with no need for mechanical parts.